Ultrasound Artefacts: What They Are and What They Tell You

An artefact is anything on the screen that does not match the anatomy. Some hide disease, some imitate it, and some — the shadow behind a stone, B-lines in the lung — are the diagnosis. Why the machine makes them, how to recognise each one, and how presets, harmonics and compounding change what you see.

Also called Ultrasound artefacts, Ultrasound artifacts, Image artefacts

Published:

Last updated:

Why artefacts happen

The machine builds the image on a few assumptions: sound travels in a straight line at 1,540 m/s through all tissue, each echo comes from a single reflection, and sound weakens at a steady rate with depth. When an assumption fails, the image shows something that is not there, puts something in the wrong place, or hides it.

AssumptionWhen it failsArtefacts it causes
Sound travels in a straight, narrow beamSound bends at tissue boundaries, and weaker side beams leave the probeRefraction duplication, edge shadowing, side lobes
Each echo comes from one reflectionSound bounces between strong reflectors before returningReverberation, comet tail, ring-down, mirror image
Sound weakens at the same rate everywhereSome tissues absorb much more or much less than averageAcoustic shadowing, posterior enhancement
Speed is 1,540 m/s everywhereSound is slower in fat and faster in some other tissuesSpeed displacement: structures drawn slightly too deep or too shallow
The beam has no thicknessThe beam has a real slice thicknessPartial volume: false echoes inside cysts and the bladder

Shadowing and enhancement

Acoustic shadowing

A dark band behind something that reflects or absorbs most of the sound — gallstones, kidney stones, bone, calcification. A bright focus with a clean, dark shadow in the gallbladder supports a stone.

Dirty shadowing

Gas reflects nearly all sound, giving a grey, streaky shadow with reverberation in it. It is typical of bowel gas and helps tell gas from a stone.

Posterior acoustic enhancement

A brighter area behind fluid — cysts, the gallbladder, a full bladder — because fluid absorbs less sound than the machine expects. It helps separate a simple cyst from a solid lump, though some solid lesions also enhance.

Edge shadowing

Thin dark lines falling from the curved edges of round structures such as cysts, the fetal head or the testis, caused by sound bending at the edge. It is not a stone.

Reverberation, comet tail and ring-down

Reverberation

Evenly spaced bright lines repeating deeper and deeper, when sound bounces between the probe and a strong reflector. A-lines in the lung are reverberations of the pleural line; the same effect puts haze in the front of the bladder.

A-lines and B-lines in lung ultrasound

Comet tail

A short, tapering trail of closely spaced echoes behind a tiny strong reflector, such as metal, cholesterol crystals in the gallbladder wall or colloid in a thyroid nodule.

Comet tails in thyroid TI-RADS

Ring-down

A long, unbroken bright band behind small gas bubbles, from fluid trapped between them resonating. Often seen with gas in bowel or in an abscess.

Mirror images, side lobes and misplaced echoes

Mirror image

A strong curved reflector — usually the diaphragm — acts like a mirror, so the liver or spleen appears again above it. This is normal. Its loss, with dark fluid above the diaphragm, suggests a pleural effusion.

The chest views in eFAST

Side lobes

Weaker beams to the side of the main beam pick up strong reflectors and paint false echoes into fluid — a fake layer of sludge in the gallbladder or debris in the bladder. They change or vanish when the probe angle changes.

Slice thickness (partial volume)

The beam has thickness, so tissue next to a small cyst is averaged into it and looks like debris. Scan in two planes and narrow the focus.

Refraction duplication

Sound bending through the muscles of the abdominal wall can double a structure in the midline — a known cause of a false “twin” gestational sac in early pregnancy.

Speed displacement

A step in a straight line, such as a needle or the diaphragm, where it passes behind fat.

Anisotropy in tendons, ligaments and nerves

Tendons and ligaments — and, to a lesser degree, nerves and muscle — look bright only when the beam meets them at a right angle. Tilt the probe a few degrees and a normal tendon turns dark, imitating a tear or tendinopathy.

  • Rock or tilt the probe until the fibres light up before calling anything abnormal.
  • Compare with the same structure on the other side.
  • Use anisotropy on purpose: tilting makes a tendon go dark against bright fat, which helps pick it out from the tissue around it.

Doppler artefacts

Aliasing

Colour wraps from red to blue, or a spectral trace wraps around, when flow is faster than the scale can show. Raise the scale or move the baseline.

Twinkling

A rapidly changing mix of colour behind a rough, strong reflector such as a kidney stone. It can help confirm a small stone.

Flash and blooming

Colour from patient or probe movement, and colour spilling outside the vessel when colour gain is set too high.

Colour and spectral Doppler

How settings change artefacts

SettingEffect on artefactsWhen to change it
Tissue harmonic imagingCuts reverberation, side-lobe clutter and haze in fluid; sharpens cyst wallsKeep on for most abdominal and cardiac work; some lung protocols turn it off to see A-lines and B-lines as they are
Spatial compound imagingAverages several beam angles: smoother image and less speckle, but weaker shadowing and enhancementTurn it off to look for the shadow behind a small stone or calcification
Speckle reduction and smoothingLess grain, but fine echoes and faint B-lines can blurLower it for lung and fine detail
Gain and TGCToo much gain fills cysts with false echoes; badly set TGC can fake or hide enhancementSet gain so fluid is black, then adjust TGC depth by depth
Focus and frequencyThe beam is narrowest at the focus; higher frequency gives a thinner slice and less partial volumePut the focus at the structure you are measuring
Probe angle and windowMoves mirror images, refraction, side lobes and anisotropyReal anatomy stays in two planes; most artefacts move or disappear

Frequently asked questions

What is posterior acoustic shadowing?
A dark band behind a structure that blocks most of the sound, such as a gallstone, kidney stone, bone or calcification. Compound imaging can weaken the shadow, so turn it off when you are looking for a small stone.
What causes reverberation artefact?
Sound bouncing back and forth between two strong reflectors, often the probe face and a bright interface such as the pleura, so the machine draws repeated, evenly spaced bright lines deeper than the real one.
What is anisotropy in ultrasound?
Tendons, ligaments and nerves look bright only when the beam strikes them at a right angle. Tilting the probe makes a normal tendon look dark, which can imitate a tear, so adjust the angle before calling it abnormal.
Are ultrasound artefacts always a problem?
No. Many are diagnostic: shadowing behind gallstones, enhancement behind simple cysts, A-lines and B-lines in the lung, and twinkling behind kidney stones. The skill is knowing which to keep and which to remove.